The incoming line reactor for medical CT machine is placed in front of the isolation transformer to improve the power factor;


Effectively limit the current surge caused by sudden change of grid voltage and operating voltage;


Smoothing the spikes contained in the supply voltage;

Effectively improve transformer power factor.

Features

1、Low loss,low temperature rise,and the operation is safe and reliable;


2、Comply with relevant IEC6060-1-1 Medical safety standards;


3、Comply with relevant IEC6060-1-1 Medical EMC standards;


4、Excellent overload capacity,and low noise;


5、Suitable for wide range of voltage input,strong anti-saturation ability.

Application Environment

1、Altitude: no more than 2000m;

2、 Operating ambient temperature: -10℃~+40℃;

3、Relative humidity:<95%.

4、Maximum temperature gradient 6K/h;

5、Maximum absolute humidity 30 g/m³;

6、Atmospheric pressure range:800 to 1060  hPa。


Technical Parameters

1、 Rated voltage: 3Φ480V;

2、Rated frequency: 50Hz/60Hz;

3、 Rated current: 20A~500A;

4、Insulation class: Class H ;

5、Protection grade: IP00.